ICS-90 Installation Instructions
28
Doc. 031856-03 12/06
a.
For flow rates less than 2 mL/min, verify that two coils are installed.
b. For flow rates of 2 mL/min or more, remove one of the backpressure
coils.
NOTE Refer to the flier Backpressure Coil Pressure Test for
Dionex Suppressors (Document No. 031759)
(shipped with the suppressor) for details about
suppressor operating pressure requirements.
2.10 Pressurizing the Eluent Reservoir
NOTE Do not pressurize the regenerant reservoir.
1. Pull out the air regulator knob and turn it fully counterclockwise to ensure
there is no pressure to the reservoir when the gas source is turned on.
2. Turn on the gas source.
3. Pull out the air regulator knob and turn it clockwise to pressurize the eluent
reservoir.
4. Adjust the pressure to between 30 and 40 kPa (5 and 6 psi). Push the regulator
knob back in.
Do not pressurize the eluent reservoir above 70 kPa (10 psi).
Ne mettez jamais les réservoirs d'éluants sous une pression
supérieure à 0,07 MPa (10 lb/po²).
Setzen Sie den Eluentbehälter auf keinen Fall einem Druck über
0,07 MPa aus.