MENU ACCESS:
SAVING THE PARAMETERS:
Press the
key during the start-
up procedure.
Press the
key several times, until the display shows
SAVE?
. Press the
key to confirm.
Complete menu on pages
24 - 25
USER_MAN_ENG_DGT100_V8
5.
Acquire the calibration points:
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Press and enter the value of the first
sample weight using the , and keys.
To confirm press
Load the scale with the second sample
weight and press
Press and select the number of calibration
points using the and keys (from 1 to 3).
To confirm press
Press and enter the value of the third
sample weight using the , and keys.
To confirm press
Unload the scale.
To acquire the zero point press
Press and enter the value of the second
sample weight using the , and keys.
To confirm press
Load the scale with the third sample weight
and press
press
For successful calibration, the value of the largest sample weight must be at least 50% of the capacity.