Technical features
POWER SUPPLY
12 - 24 Vdc LPS or with class 2 power supply.
MAXIMUM ABSORPTION
(without load cells)
DGT100: 100 mA at 12 V / 70 mA at 24 V (2 W);
DGT100AN: 185 mA at 12 V / 90 mA at 24 V (2.5 W);
DGT100PB: 410 mA at 12 V; 220 mA at 24 V (5 W).
OPERATING TEMPERATURE
From -15°C to +40°C.
DISPLAY DIVISIONS
10000e, 2 x 3000e for legal weighing, expandable up to 800,000 for internal use (with a
minimum cell signal of 1.6 mV/V).
SAMPLING SPEED
Up to 400 conv. / sec.
MINIMUM VOLTAGE PER DIVISION
0.3 µV
(approved instrument)
; 0.03 µV
(non-approved instrument)
.
COUNTING RESOLUTION
1,500,000 points (with input signal 3 mV/V).
DISPLAY
6-digit, 100 mm high, with very efficient SMD LEDs (software adjustment of the brightness
intensity); visible from over 25 m, with viewing angle up to 120° and with power up to 750
mcd.
SIGNALS
6 status indicator LED lights.
KEYPAD
5 keys.
TARE FUNCTION
Subtraction possible over the entire range.
LOAD CELL POWER SUPPLY
5 Vdc ± 5%, 120 mA (max 8 cells of 350 Ω).
LOAD CELL CONNECTION
6 wires (CELL1) with sense, 4 wires (CELLS 2, 3, 4).
CONNECTABLE CELLS
Up to 8 350 Ω cells.
CASE
IP68 AISI 304 stainless steel, complete of wall fixing bracket.
SERIAL OUTPUTS
1 RS485 bidirectional port;
1 configurable RS232 bidirectional port for connection to printer;
1 PROFIBUS port (DGT100PB* version);
*
Fieldbus models are not equipped with port 485.
OUTPUTS / INPUTS
2 fotomosfet outputs NO or NC, configurable as programmable weight thresholds:
48 Vac 0.15 A max (or 60 Vdc 0.15 A max);
2 configurable inputs (optocouplers): 12 - 48 Vdc;
Input reading and output update time: 1 msec;
Opto-isolated analog output with 16 bits, optionally 4 - 20 mA, 0 - 5 Vdc or 0 - 10 Vdc (DG-
T100AN version). The maximum applicable resistance on the current output is 350 Ω while
the minimum applicable resistance on the voltage output is 10 kΩ.
LOAD CELL SENSITIVITY
Maximum sensitivity of the connectable load cells: 6 mV/V.
FIELDBUS UPDATE RATES
Up to 16 Hz.