33
Optimized layout for A4 print.
i
TYPE
A
888888
Setup
C
ConfiG
1
Stabil
Param
1
5
TYPE
A
888888
Setup
C
ConfiG
1
Stabil
div.stb
Param
1
5
5
LEGEND:
Indicates repeated pressing of
the key.
Parameter visible only under
certain conditions.
Parameter or menu subject to
approval.
Default value of the parameter.
USER_MAN_ENG_DGT100_V8
In the case of an approved instrument, it is possible to select only FLT
filters
.
Filter and stability
Filter
Response time (ms)
Updating frequency
(Hz)
Filter
Response time (ms)
Updating frequency
(Hz)
h.r.0
1230
6
flt 2
590
25
h.r.1
1435
6
flt 3
980
25
h.r.2
1840
6
f.f.50.1
440
50
h.r.3
2050
6
f.f.50.2
390
50
h.r.4
3280
6
f.f.50.3
340
50
h.r.5
4100
6
f.f.100.1
110
100
h.r.6
4920
6
f.f.100.2
210
100
dyn.0
1640
6
f.f.100.3
260
100
dyn.1
2050
6
f.f.100.4
280
100
dyn.2
2460
6
F.f.200.1
170
200
dyn.3
2460
6
f.f.200.2
155
200
flt 0
250
25
f.f.200.3
140
200
flt 1
390
25
f.f.400
50
400
Filter adjustment
The value 0 disables the stability control.
By entering a value other than 0, you enable stability control.
Enter the number of deviation divisions beyond which the indicator detects instability.
It is possible to decide that tare, zero and print functions (from keypad or serial command / PLC) are performed only if the weight is stable.
From 0 to 99.
= 02.
= 02.
Stability detection sensitivity
Содержание DGT100