56
Optimized layout for A4 print.
s.1 on
0.000
s.1 off
s.2 on
s.2 off
inputs
output
an.out
13
12
14
Cts.st.
11
ser
10
USER_MAN_ENG_DGT100_V8
Programming the Setpoints
In weighing mode, if the output functions (
1 Gross
/
2 net
) have been set correctly, pressing
for 3 seconds
will enter the setpoint
programming menu:
Once you have entered the desired values, press . The display shows “
store
” and returns to weighing mode.
visible only if
rl.ist
=
ist.on
Output 1 deactivation value.
Output 2 deactivation value.
Output 1 activation value.
Output 2 activation value.
visible only if
rl.ist
=
ist.on
Inputs
Outputs
Analog output
Checking the status of the inputs:
value 0 indicates that the input is disabled, value 1 indicates that the input is enabled.
Usethe and keys to display the two inputs.
Activation of the output shown on the display (
rel.1
/
rel.2
).
Use the and keys to activate the two outputs.
Analog output test.
Use the , , keys to enter the D/A point value of the analog output.
Press the key to confirm and update the V / mA value of the analog output.
Checking the CTS signal of the printer (on) connected to the PRN port.
CTS signal
Bridge between serial ports (for manufacturer’s use).
Serial ports