CLM–70
©
Dinel, s.r.o.
12
•
The level meter of the CLM-70_-
00 type is provided without a
measuring electrode. Therefore,
the measuring electrode is nec
-
essary to be made and mounted
to the level meter body. The elec
-
trode diameter must be between
8–10 mm. For connection, the
electrode must have an M8 male
thread. For the connection proce-
dure, see chapter 12, p. 22.
Fig. 10: The level meter without an electrode
For the level meter of the CLM-70_-00 type, the manufacturer is not responsible for
defects related to the connected measuring electrode!
It is intended for non-conductive media only, as only a non-insulated rod or rope
can be connected.
VARIANT WITHOUT AN ELECTRODE
APPLIES TO: CLM-70_-00
• It is advisable to run the cable un
-
der a cable gland (sagging down
-
wards). This prevents possible
ingress of moisture through the
cable gland
. Rain and condensing
water can flow freely downwards,
see Fig. 9.
• The cable gland and the top lid
must be
sufficiently tightened
to prevent moisture ingress.
Fig. 9: Preventing moisture from ingress
through the cable gland
MOISTURE
APPLIES TO: All types