4–16
Aligning the Video Monitor
4.7 Using the Radiance Meter
Use the radiance meter as follows:
1.
Remove the cap from the radiance meter sensor head and connect the
occluder to the radiance meter (Figure 4–12).
CAUTION
Avoid excessive force when tightening the occluder or you may
damage the radiance meter. Do not touch the exposed filter
after the protective cap is removed.
O c c u l u d e r
M A - 0 6 4 7 - 9 0 . D G
A C A d a p t e r
C o n n e c t o r
S e n s o r
H e a d
P o w e r
S w i t c h
R a n g e
S w i t c h
A C A d a p t e r
Figure 4–12
Radiance Meter
2.
Connect the ac line adapter to the radiance meter and plug it into a
wall outlet.
NOTE
Be sure that your radiance meter is in calibration.