V3.01.51
4
Receiver (WR420/425):
WR420 features a WR400 Receiver and WiZARD2.4 Software. WR425
features a WR400 Receiver and WiZARD2.4 SECURE 21CFR11
Compliant Software.
The Receiver forwards data packets from the Loggers to the WiZARD2.4
database.
Must always be on and connected to the PC via a USB cable. It should be
powered via the AC adapter and have Backup Batteries installed.
Receivers do not require calibration.
Logger (WT400/420/440/WH420/425/445):
Take readings at regular intervals and send those readings to the
Receiver at regular transmission intervals.
Loggers are setup via USB at the host PC.
Have a practical unobstructed operating range of up to 300 ft (150 ft.
standard) from Receiver. Any metal infrastructure or other wireless
devices will significantly reduce operating range.
Will save logged data when communications with the Receiver have been
interrupted. Once communications are restored, saved logged data will be
sent.
Loggers should be calibrated every 6 to 12 months.
Repeater (WA400):
Extends the distance of a Logger from the Receiver by forwarding the data
packet from the Logger to the Receiver.
Must be setup in the WiZARD2.4 Software in order to operate.
Logger(s) must be told to send data packet to the Repeater not the
Receiver.
Repeater can forward data packets for multiple Loggers.
Repeaters can forward data packets to another Repeater.
Repeaters do not require calibration.
Signal Sensor (WS400):
Allows the user to determine the signal strength of the Receiver, and any
setup Repeaters, at a current or potential Logger or Repeater location.
Does not look for Loggers only Repeaters and Receivers