8 Interface Specifications
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 74 of 81
Correlation Flag
Correlation Type
Flag Code
Default
D
User Correlation
U
MGL Correlation (for
Ca)
M
Flame Correlation
F
Error Flag
Correlation Type
Flag Code
MV Range Error
V
Noise Error
N
Drift Error
D
Range Low
L
Range High
H
Too Low
<
Too High
>
Not Calibrated
X
Not Consistent
!
Cannot Measure
@
Li Cannot Measure
M
Reading 1 mV Sign
This character represents if value is positive or negative. “+” represents positive and “-” negative
numbers. During calibration, this represents STD-A. During sample analysis, it represents
Sample mV.
Reading 1 mV Value
Ion concentration value is sent out in 7 Digits. This number is 1000 times the concentration
determined in the analysis. For example, 01234560 represent value of 1234.56. During
calibration, this represents STD-A. During sample analysis, it represents Sample mV.
Reading 1 mV Try Count (
Valid only for SMARTLYTE Calibration
)
The number represents number of tries instrument ran the standard fluid before producing stable
results. During calibration, this represents STD-A.
Reading 2 mV Sign
This character represents if value is positive or negative. “+” represents positive and “-” negative
numbers. During calibration, this represents STD-B. During sample analysis, it represents
Standard Fluid mV.
Reading 2 mV Value
Ion concentration value is sent out in 7 Digits. This number is 1000 times the concentration
determined in the analysis. For example, 01234560 represent value of 1234.56. During
calibration, this represents STD-B. During sample analysis, it represents Standard Fluid mV.
Содержание SmartLyte
Страница 1: ......
Страница 11: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 11 of 81 1 OVERALL FLOW ...
Страница 12: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 12 of 81 2 MAIN MENU HUMAN ...
Страница 13: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 13 of 81 3 MEASUREMENT SEQUENCE ...
Страница 17: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 17 of 81 6 MAINTENANCE ...
Страница 22: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 22 of 81 9 SERVICE MENU ...
Страница 25: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 25 of 81 11 POWER UP ...
Страница 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Страница 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Страница 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Страница 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Страница 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Страница 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Страница 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...
Страница 81: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 81 of 81 9 3 Display 9 4 Valve ...