5 Maintenance and Troubleshooting
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 53 of 81
5.2.
Error Messages and Troubleshooting
STATUS: INVALID CALIBRATION
This message is only displayed when a calibration has been interrupted by pressing NO,
opening the sample door, or none of the electrodes are calibrated. Perform a calibration to
return to the H-READY Mode.
STD A NOT FOUND
This message is displayed when the sample sensor is unable to properly detect the STD A
solution in a programmed time period. As the analyzer draws a sample of STD A, the
leading edge of the solution is detected by the sample sensor. The sample sensor must
also detect a continuous flow of STD A Solution into the sample chamber, lasting several
seconds, without encountering air bubbles. After several seconds, the trailing edge of the
STD A sample is sensed, the peristaltic pump stops and measurement occurs.
Possible causes and remedies:
•
Check for air leaks preventing the Standard from being drawn into the chamber.
Ensure that all O-rings are in place. There must be a continuous draw of standard
solution, free of air bubbles. Check solenoid V for proper sealing of the vent line.
Check fill port to ensure the absence of air leaks.
•
Check for clots or crystals formed in the STD A tubing or electrode chamber.
•
Ensure that the sample sensor is plugged in securely and perform Check Sample
Sensor to ensure sensor is operating correctly. Clean, adjust or replace sample sensor
to correct sample sensor operation.
•
Replace peristaltic pump tube set to ensure proper sample aspiration.
•
Check the fluid remaining in the Fluid Pack. If less than 5% remains, replace the
Fluid Pack.
•
Prime system; look for fluid flow through system.
STD B NOT FOUND
STD C NOT FOUND
The system for drawing STD B or C is like that of STD A. If STD A is drawn properly
and STD B or C is not detected, check STD B tubing for crystallization. Perform checks
for STD A detection problems to correct. During the calibration sequence, STD B or C is
drawn into the measuring chamber prior to STD A, therefore check if STD A calibrant
can be properly detected to isolate possible fluidic faults. If neither STD A, B nor C are
detected check for air leaks or blockages within the fluidic path.
Содержание SmartLyte
Страница 1: ......
Страница 11: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 11 of 81 1 OVERALL FLOW ...
Страница 12: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 12 of 81 2 MAIN MENU HUMAN ...
Страница 13: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 13 of 81 3 MEASUREMENT SEQUENCE ...
Страница 17: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 17 of 81 6 MAINTENANCE ...
Страница 22: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 22 of 81 9 SERVICE MENU ...
Страница 25: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 25 of 81 11 POWER UP ...
Страница 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Страница 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Страница 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Страница 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Страница 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Страница 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Страница 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...
Страница 81: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 81 of 81 9 3 Display 9 4 Valve ...