5 Maintenance and Troubleshooting
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 56 of 81
CONDITIONER NOT FOUND
When the conditioning cycle has been initiated, the analyzer checks for proper aspiration
of the conditioning solution using the sample sensor. If the solution is not detected in the
proper time, this error message will be displayed.
Possible causes and remedies:
•
Ensure that a sufficient volume of conditioning solution is presented for aspiration and no air
bubbles are present as solution is aspirated.
•
Check to ensure that the solution is being properly aspirated through the sampling
mechanism. The sample path must be free of leaks or clots.
•
Verify that sample sensor is correctly plugged in and perform the CHECK SAMPLE SENSOR
to verify its proper operation.
•
Check pump tubing and replace as necessary to ensure correct pump rate.
INTERFACE CHECK ERROR
This message is displayed only during the ACTIVATE INTERFACE. Pins 2 and 3 must
be shorted together to perform the ACTIVATE INTERFACE. Make sure that the pins are
not shorted to chassis ground.
Possible cause and remedy:
•
Call Product Support
PRINTER JAM OR DEFECT
This message is displayed when the printer head is jammed and tries to print. It is a
temporary message that is displayed for 2 seconds and then sample results are displayed.
To clear paper jam remove printer from the analyzer and remove jammed paper. Replace
printer and retry. The printer can be removed by grasping the rear edge of the printer
assembly and pulling it to the front. The printer can be removed with the Analyzer power
on. To free up jammed paper, turn the small gears on the left side of the printer module.
Содержание SmartLyte
Страница 1: ......
Страница 11: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 11 of 81 1 OVERALL FLOW ...
Страница 12: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 12 of 81 2 MAIN MENU HUMAN ...
Страница 13: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 13 of 81 3 MEASUREMENT SEQUENCE ...
Страница 17: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 17 of 81 6 MAINTENANCE ...
Страница 22: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 22 of 81 9 SERVICE MENU ...
Страница 25: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 25 of 81 11 POWER UP ...
Страница 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Страница 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Страница 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Страница 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Страница 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Страница 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Страница 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...
Страница 81: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 81 of 81 9 3 Display 9 4 Valve ...