5 Maintenance and Troubleshooting
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 55 of 81
SHUT DOOR
This message occurs when either of two conditions exists.
1.
Sample door is not closed within 20 seconds after samples are in place.
2.
Sample door has been opened and no sample is detected within 20 seconds.
Possible causes and remedies:
•
Close sample door within the time allowed.
•
Check needle mechanism light gate.
•
Check whether sample sensor is plugged in.
SAMPLE NOT FOUND
This message is displayed when any of the following conditions occur:
1.
During sampling, either a bubble in the sample is detected or the sample size is smaller than
the minimum sample required.
2.
The sample door is opened and no sample is fed into the analyzer.
3.
Sample door is not closed within 20 seconds after sample is detected.
Possible causes and remedies:
•
Check whether the sample sensor is plugged in and perform CHECK SAMPLE
SENSOR to verify that it operates correctly.
•
Check sample draw and look for clots or leaks in the sampling system.
CLEANING SOLUTION NOT FOUND
When the cleaning cycle has been initiated, the analyzer checks for proper aspiration of
the cleaning solution using the sample sensor. If solution is not detected in the proper
time, above error message will be displayed.
Possible causes and remedies:
•
Ensure that a sufficient volume of cleaning solution is presented for aspiration and no air
bubbles are present as solution is aspirated.
•
Check to ensure that the solution is being aspirated through the sampling mechanism
correctly. The sample path must be free of leaks or clots.
•
Verify that sample sensor is properly plugged in and perform the CHECK SAMPLE
SENSOR to verify the correct operation of the sample sensor.
•
Check pump tubing and replace as necessary to ensure a correct pump rate.
Содержание SmartLyte
Страница 1: ......
Страница 11: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 11 of 81 1 OVERALL FLOW ...
Страница 12: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 12 of 81 2 MAIN MENU HUMAN ...
Страница 13: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 13 of 81 3 MEASUREMENT SEQUENCE ...
Страница 17: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 17 of 81 6 MAINTENANCE ...
Страница 22: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 22 of 81 9 SERVICE MENU ...
Страница 25: ...3 Software Operation ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 25 of 81 11 POWER UP ...
Страница 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Страница 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Страница 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Страница 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Страница 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Страница 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Страница 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...
Страница 81: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 81 of 81 9 3 Display 9 4 Valve ...