7. Fault Finding

The 

VENTURI

 MINO power unit features audible and graphic indicators to alert when user 

intervention is required. Once corrective action has been taken the alert will self-cancel and normal 
operation will resume. The display screen will toggle between the alert graphic and the instruction to 
refer to the user manual for more information (Fig. 4). The illuminated display and sounder will persist 
whilst the alert state is active (NB. Tilt indicator is a graphic alert only, not audible). Should an alert 
persist or immediately re-occur, contact Direct Healthcare Group. All sounders can be silenced and 
messages cleared by pressing and holding the RUN/STOP/UNLOCK button to switch the power unit off, 
whilst corrective action is taking place. The time period before the alert is indicated is shown in italics.

NO CANISTER

 (Fig. 5) (

immediate

) – indicates canister is missing or is not correctly fitted. The power unit will fail 

to operate whilst this message is displayed. Check that canister is correctly located and secured, as detailed on page 5.

CANISTER FULL

 (Fig. 6) (

60 secs

) – indicates that the canister is full and should be removed/replaced, as detailed 

on page 6. The power unit will cease to run until corrective action has been taken.

CONSTRICTED TUBE/LOW FLOW*

 (Fig. 7) (

31 mins

) – indicates tube blockage/constriction or state of low 

flow. Check that low flow is not caused by pinched or bent tubing. If exudate is pooling in the wound bed, the dressing 
should be changed. This alert will self-cancel after 5 minutes and will re-occur if the condition persists. NB. This alert 
may arise under normal operation if the wound exudate levels are very low.

D H G - H E A LT H C A R E . C O M

9

U S E R   M A N U A L

!

Fig. 4

!

!

Fig. 10

80

-mmHg

Fig. 11

Fig. 12

E1

!

?

!

Fig. 8

!

!

Fig. 9

+

!

!

Fig. 5

Fig. 6

!

!

!

!

!

Fig. 7

CHECK DRESSING* 

(Fig. 8)

 / DISCONNECTED TUBE*

 (Fig. 9) (

6 mins

) – indicates vacuum pressure has 

fallen below minimum allowable levels, due to either a leak in the dressing or a tube disconnection. The power unit 
will continue to run whilst this message is displayed. Check that wound dressing is completely sealed and that all 
tubing connections are secure..

TILT

 (Fig. 10) (graphic alert only) (

30 secs

) – activates when the power unit is placed at an angle that could affect 

the canister full indication. The power unit will continue to run whilst this message is displayed. Return the power unit 
to an upright position.

LOW BATTERY

 (Fig. 11) (only appears during battery operation) – battery icon at the top left of the normal 

operation display screen flashes on/off, indicating battery charge is low and battery needs recharging. Plug into a 
power source to charge.

FATAL ERROR

 (Fig. 12) – the screen displays a letter E followed by a number, indicating an irrepairable fault. This 

screen should never appear during normal operation but if it does, please contact Direct Healthcare Group.
* the sounder can be silenced temporarily by briefly pressing the UP/DOWN button.

Содержание TG600/14

Страница 1: ...DHG HEALTHC AR E C O M DHG HE ALTHCAR E C O M VENTURI MINO User Manual...

Страница 2: ...VENTURI MINO DHG HEALTHCARE COM...

Страница 3: ...tions and Information 4 4 How to Apply NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 Battery Information 7 6 Care and Main...

Страница 4: ...n Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with...

Страница 5: ...gery Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabetic neuropathic foot ulcers venous leg ulcers post op surgical wounds including flaps a...

Страница 6: ...near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI MINO TG600 14 vacuum power unit...

Страница 7: ...o the VENTURI MINO power unit by lining up the two parts pushing together and twisting to lock Fig 1 Remove the bung from the connector port this can be retained to cap the used canister prior to disp...

Страница 8: ...ubing as above this can be reconnected to new canister and unclamped if wound dressing is not being changed Rotate canister to unlock and remove from power unit Dispose of used canister according to l...

Страница 9: ...begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen 5 3 Battery Information A fully charged battery should operate the power unit continuously fo...

Страница 10: ...ocial care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used...

Страница 11: ...heck that low flow is not caused by pinched or bent tubing If exudate is pooling in the wound bed the dressing should be changed This alert will self cancel after 5 minutes and will re occur if the co...

Страница 12: ...above mains adapter is considered part of the ME equipment The VENTURI MINO power unit must only be used with the specific external power adapter as supplied by Direct Healthcare Group EXPECTED SERVIC...

Страница 13: ...ntinuous resets Try to ascertain the source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to c...

Страница 14: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the V...

Страница 15: ...rs such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electroma...

Страница 16: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...