NB. The power unit will automatically lock 1 minute after last button operation when running to prevent inadvertent 
operation of button functions as indicated by the lock symbol     on the display screen. Press and hold the RUN/
STOP/UNLOCK button until the lock symbol disappears fom the screen if further button operation is needed (i.e. 
change of vacuum level, or to switch off the power unit). The power unit will lock again 1 minute after last button 
operation. Pressing the RUN/STOP/UNLOCK button will also illuminate the display screen for 10 seconds.

4.  Once the power unit is running, observe the wound site. The dressing should contract noticeably, becoming firm 

to the touch. If the dressing fails to contract, the dressing has not been completely sealed. Reinforce the dressing 
seal and/or adjust the drain and initiate suction again.

 

NB. During system set-up, various checks and internal tests take place and if any intervention is required an alert 
will occur, but not until several minutes after set-up.

 

WARNING: Particularly when used outside of a medical institution, get immediate medical assistance from 
those responsible for the prescription and setting of the NPWT system should any of the following occur:- 

obvious bleeding or pain; the wound site or exudate presents unexpected changes in its condition, colour or odour; 
the wound dressing becomes detached or ineffective; the tubing becomes blocked.

5.  To change or remove dressing, unlock power unit (press and hold the RUN/STOP/UNLOCK button for 3 seconds 

until lock symbol disappears), and switch off power unit (press and hold the RUN/STOP/UNLOCK button until power 
unit beeps). Clamp the drain tubing and remove by lifting the tubing up from the routing channel and pulling from 
the tubing connector on the canister. Dispose of used Wound Care Set according to local clinical waste policy. If 
required, apply new Wound Care Set and continue NPWT.

6.  Canisters should be replaced as required or weekly. To change canister, clamp and remove Wound Care Set tubing 

as above (this can be reconnected to new canister and unclamped if wound dressing is not being changed). Rotate 
canister to unlock and remove from power unit. Dispose of used canister according to local clinical waste policy. 
If continuing NPWT, attach new canister and connect Wound Care Set tubing as previously described.

7.  To switch off the power unit, unlock (press and hold the RUN/STOP/UNLOCK button for 3 seconds until the lock 

symbol disappears) release key and press and hold the RUN/STOP/UNLOCK button again until power unit beeps.

8.  Place the user manual in a safe place for future use.

5. Operation Guidelines

5.1 User Information

The 

VENTURI

 MINO power unit will automatically lock 1 minute after last button operation when running to prevent 

inadvertent operation of button functions as indicated by the lock symbol     on the display screen. Press and hold 
the RUN/STOP/UNLOCK button until the lock symbol disappears fom the screen if further button operation is needed 
(i.e. change of vacuum level, or to switch off the power unit). The power unit will lock again 1 minute after last button 
operation. Pressing the RUN/STOP/UNLOCK button will also illuminate the display screen for 10 seconds.
The 

VENTURI

 MINO power unit is suitable for ambulatory use at the advice of a Physician and is supplied with a 

carry bag which can be worn on the shoulder, across the body or around the waist. When placing the power unit in the 
carry bag, make sure that it is oriented so that the tubing exits the top of the bag without kinking, bending or straining 
against the connector port to ensure correct operation.
To ensure correct operation, the 

VENTURI

 MINO power unit should always be kept in an upright position. The power 

unit will display an alert if the unit is over tilted (refer to Fault Finding on page 9).

Please take care when transporting the device not to drop the system onto a hard surface, bend the canister 
connector or pull on the dressing tube excessively. If the connector is broken a visual and audible alarm will 

be activated and the canister connector should be replaced. Due to contamination the broken connector should be 
disposed of as clinical waste.

D H G - H E A LT H C A R E . C O M

6

V E N T U R I

®

  M I N O

Содержание TG600/14

Страница 1: ...DHG HEALTHC AR E C O M DHG HE ALTHCAR E C O M VENTURI MINO User Manual...

Страница 2: ...VENTURI MINO DHG HEALTHCARE COM...

Страница 3: ...tions and Information 4 4 How to Apply NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 Battery Information 7 6 Care and Main...

Страница 4: ...n Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with...

Страница 5: ...gery Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabetic neuropathic foot ulcers venous leg ulcers post op surgical wounds including flaps a...

Страница 6: ...near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI MINO TG600 14 vacuum power unit...

Страница 7: ...o the VENTURI MINO power unit by lining up the two parts pushing together and twisting to lock Fig 1 Remove the bung from the connector port this can be retained to cap the used canister prior to disp...

Страница 8: ...ubing as above this can be reconnected to new canister and unclamped if wound dressing is not being changed Rotate canister to unlock and remove from power unit Dispose of used canister according to l...

Страница 9: ...begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen 5 3 Battery Information A fully charged battery should operate the power unit continuously fo...

Страница 10: ...ocial care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used...

Страница 11: ...heck that low flow is not caused by pinched or bent tubing If exudate is pooling in the wound bed the dressing should be changed This alert will self cancel after 5 minutes and will re occur if the co...

Страница 12: ...above mains adapter is considered part of the ME equipment The VENTURI MINO power unit must only be used with the specific external power adapter as supplied by Direct Healthcare Group EXPECTED SERVIC...

Страница 13: ...ntinuous resets Try to ascertain the source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to c...

Страница 14: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the V...

Страница 15: ...rs such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electroma...

Страница 16: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Отзывы: