background image

D H G - H E A LT H C A R E . C O M

1 2

V E N T U R I

®

  M I N O

Declaration – Electromagnetic Emissions

Declaration – Electromagnetic Immunity

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)

Immunity Test

IEC 60601 Test 

Level

Compliance Level

Electromagnetic

Environment - Guidance

Electrostatic discharge (ESD) 
IEC 61000-4-2

± 8kV contact
± 15kV air

± 8kV contact
± 15kV air

The relative humidity should be 
at least 5%.

Electrical fast transient/burst 
IEC 61000-4-4

± 2 kV For mains 
supply lines
100kHz repetition 
frequency

± 2 kV For mains supply 
lines
100 kHz repetition 
frequency

Mains supply quality for the 
mains adapter should be that 
of a typical commercial and/or 
hospital environment.

Surge
IEC61000-4-5

± 2kV Line(s) to 
ground
± 1kV line(s) to line

± 2kV Line(s) to ground
± 1kV line(s) to line

Mains supply quality for the 
mains adapter should be that 
of a typical commercial and/or 
hospital environment.

Voltage dips, short interruptions 
and voltage variations on mains 
supply IEC 61000-4-11

0 % U

T

; 0,5 cycle 

At 0°, 45°, 90°, 
135°, 180°, 225°, 
270° and 315°

0 % U

T

; 0,5 cycle 

At 0°, 45°, 90°, 135°, 
180°, 225°, 270° and 
315°

Mains supply quality for the 
mains adapter should be
that of a typical commercial 
and/or hospital environment.
In the event of a mains 
interruption the VENTURI
system will automatically use 
internal battery power,
unless the battery is exhausted.

0 % U

T

; 1 

cycle 
70 % U

T

; 25/30 

cycles
Single phase: at 0°

0 % U

T

; 1 cycle 

70 % U

T

; 25/30 cycles

Single phase: at 0°

Voltage interruptions

0 % U

T

; 250/300 

cycle

0 % U

T

; 250/300 cycle

Mains frequency (50/60Hz) 
magnetic field IEC61000-4-8

30 A/m

30 A/m

Mains frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical commercial 
and/or hospital environment.

Note: U

T

 is the A.C. mains voltage prior to application of the test level.

Guidance and Manufacturer’s Declaration: Electromagnetic Emissions (IEC 60601-1-2)

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions
CISPR 11

Class B

The VENTURI systems are suitable for use in all 
establishments, including domestic establishments 
and those directly connected to the public 
low-voltage power supply network that supplies 
buildings used for domestic purposes.

Harmonics emissions
61000-3-2

Class A

Voltage fluctuations / flicker emissions
61000-3-3

Complies

Содержание TG600/14

Страница 1: ...DHG HEALTHC AR E C O M DHG HE ALTHCAR E C O M VENTURI MINO User Manual...

Страница 2: ...VENTURI MINO DHG HEALTHCARE COM...

Страница 3: ...tions and Information 4 4 How to Apply NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 Battery Information 7 6 Care and Main...

Страница 4: ...n Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with...

Страница 5: ...gery Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabetic neuropathic foot ulcers venous leg ulcers post op surgical wounds including flaps a...

Страница 6: ...near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI MINO TG600 14 vacuum power unit...

Страница 7: ...o the VENTURI MINO power unit by lining up the two parts pushing together and twisting to lock Fig 1 Remove the bung from the connector port this can be retained to cap the used canister prior to disp...

Страница 8: ...ubing as above this can be reconnected to new canister and unclamped if wound dressing is not being changed Rotate canister to unlock and remove from power unit Dispose of used canister according to l...

Страница 9: ...begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen 5 3 Battery Information A fully charged battery should operate the power unit continuously fo...

Страница 10: ...ocial care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used...

Страница 11: ...heck that low flow is not caused by pinched or bent tubing If exudate is pooling in the wound bed the dressing should be changed This alert will self cancel after 5 minutes and will re occur if the co...

Страница 12: ...above mains adapter is considered part of the ME equipment The VENTURI MINO power unit must only be used with the specific external power adapter as supplied by Direct Healthcare Group EXPECTED SERVIC...

Страница 13: ...ntinuous resets Try to ascertain the source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to c...

Страница 14: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the V...

Страница 15: ...rs such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electroma...

Страница 16: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Отзывы: