4. Maintenance
Check that
Action if a problem is noticed
Frame
There is no mechanical damage.
If mechanical damage is present, contact DHG Customer Services.
There are no scratches.
Touch up as necessary.
End plugs/lower frame fitted and not
damaged.
Fit new end plugs.
Armrest
Armrests are intact
Replace if damaged.
Armrest widening is functional.
Replace lock knob, see parts section.
Handle
Grips are not damaged.
Replace handle grips.
Handle adjustment is functional.
Adjust lock screws/plates as required.
Electronics
Check that the charging cable is con-
nected to the control box.
Fit new charging cable; this must always be connected to the control
box.
Check that the hand control is connect-
ed to the control box.
Connect or fit new hand control as required. Tighten; replace with
new fasteners as required.
Check that the battery, actuators and
control box are not loose.
If any part needs to be replaced refer to the wiring diagram in the
control box wiring section.
Height adjustment
Raising and lowering are functional.
Fit new guide sleeves (in frame) or end plugs (in chrome
tubes) as required.
The walker feels stable at maximum
height and width.
Tighten all nuts and bolts.
There is no movement between the
vertical frame and bottom frame.
Tighten all nuts and bolts.
Height adjustment clamps lock.
Replace clamps.
The electric actuator fitting must be free
from movement.
Tighten all nuts and bolts; replace bolts and nuts if necessary.
Height adjustment via hand control is
functional. The electric actuator must
run smoothly at a constant speed.
Check that the battery/hand control/ actuator are connected accord-
ing to the control box wiring section. Charge the battery. See parts
section
Castors and brakes
Castors roll easily, tread not damaged.
Replace castors.
The castors are firmly fastened to the
lower frame.
Tighten the castor bolt and nut. Use thread locker.
Castor brakes functional on all castors.
Handbrake functional.
Replace castors or handbrake assembly. Clean and adjust the brake.
The castors are sealed and it is not recommended to dismantle them;
instead, replace the whole castor assembly.
•
Before maintenance. clean walker according to cleaning instructions.
•
The device must undergo thorough inspection at least once per year. Inspection must be performed by
authorised personnel and in accordance with Direct Healthcare Group’s instructions.
•
Repairs and maintenance may only be done by authorised personnel using original spare parts.
•
Power cord may only be replaced by DHG service personnel or technical personnel trained by DHG.
•
DHG can provide, upon request neccessary technical information in order to assist maintenace and repair of
the device.