x
P - patient,
X
C - caregiver
Control box
Linak, voltage 100V/240V, Max 1.5A, 50/60 Hz, IPX6, Class II (Double insulated)
Batteries
Linak, 25,2V/2,1Ah, sealed,Li-Ion, 300mA IPX6, Cable length 200mm
Lifting Actuator
Linak, DC 24 V, 4,3A, IPX6. Duty cycle: Max 10% or 2 minutes continuous use
followed by 18 minutes not in use, Push: 1500N. Stroke 400mm +/-2 mm
Widening Actuator
Linak, DC 24 V, 3A, IPX6. Duty cycle: Max 10% or 2 minutes continuous use
followed by 18 minutes not in use, Push/Pull: 1200N. Stroke 180mm +/-2 mm
Hand Control
Cable length coiled 600 mm, IPX6
IP class
IPX5
Material
Steel
Castors
125/100/75mm
Weight
33 kg
Dimensions
For dimensions, see “7. Dimensions” page
Expected Service life
Walkers: 8 years
Textile products: 2 years under normal use. Two wash cycles a week are considered
as normal use. However, the total lifetime is dependent on the use, maintenance and
storage of the product.
Max patient weight
150 kg / 330 Ibs
Operating environment
5 °C to 40°C, relative humidity 20% to 80% non-condensing, and atmospheric
pressure 700 to 1060 hPa.
Storage and transportation
environment
5 °C to 50 °C, relative humidity 20% to 80% non-condensing, and atmospheric.
Environment pressure 700 to 1060 hPa.
Materials in contact with
human skin
Materials that patient and operator/caregiver are exposed to during usage of the
product for prolonged time:
Meditap B-365 (synthetic leather cushions) P
x
Polyurethane (on PU cushions) P
Dryflex (handles) P
PA6 + 20% GF (brake lever) P
Hand control C
X
Applied parts:
Armrest P
Handles P
Knee pad P
Foot rest support P
Foot rest support clamps C
Handles adjusting knobs C
Hand controller and its holder. C
Handles to adjust the armrest width. C
Hand control C
Charger C
Charger cable C
Footplate P
5. Technical Specifications