46
12 Measuring Ranges
12.1 Strain-Gauge Transducer (DMS)
The measurement instrument ME6000 has five measuring ranges for strain-gauge
transducers of the DEPRAG company.
All measuring ranges will be rounded up or down. The measuring ranges according to
N·cm
tric
al place.
the standard calibration method are:
Measuring Range 1:
MP2DMS
20 - 200 N·cm
Measuring Range 2:
MP7DMS 100
-
700
Measuring Range 3:
MP25DMS
5 - 25 N·m
Measuring Range 4:
MP160DMS
20 - 160 N·m
Measuring Range 5:
MP500DMS
60 - 500 N·m
12.2 Piezo Electric Transducer
The measurement instrument ME6000 has six measuring ranges for piezo elec
transducers of the DEPRAG company. Those depend on the entered calibration value
(C-Value).
The measuring ranges will be rounded up or down to the first decim
Measuring Range 1:
20 pC
Measuring Range 2: 100 pC
Measuring Range 3: 200 pC
Measuring Range 4: 1000 pC
Measuring Range 5: 2000 pC
Measuring Range 6: 10000 pC
Calculation lower limit:
( (Measuring Range * 5) / C-Value ) * 16%
= 0.093 N·m
Calculation upper limit:
example 1 range 1 :
( (20 pC * 5) / 172 pC/N·m ) * 0.16
0.093 N·m * 100 => 9.3 N·cm
example 2 range 4 :
( (1000 pC *5) / 172 pC/N·m ) * 0.16 ) = 4.65 N·m
( (Measuring Range * 5) / C- Value) * 96%
example 1 range 1 :
( (20 pC * 5) / 172 pC/N.m ) * 0.96 = 0.558 N·m
0.558 N·m * 100 = 55.8 N·cm
example 2 range 4 :
( (1000 pC *5) / 172 pC/N·m ) * 0.96 ) = 27.9 N·m