6
Cleaning: Automated
• May be cleaned using an ultrasonic water bath for
15 minutes.
Cleaning: Manual
• Rinse all instruments under running water to remove any
gross debris.
Disinfection:
Additional disinfection is not necessary prior to sterilization.
Packaging:
Place instrument in a paper or paper/plastic steam sterilization
pouch. Instruments may be loaded into dedicated instrument
trays. Remove Endosonic Files from PEC and PEC2 inserts
prior to packaging.
Sterilization:
Use a steam autoclave. Place bagged instruments into the
steam autoclave, paper side up when using a paper/plastic
pouch.
Gravity Steam Sterilization
Full Cycle: Bagged -135°C (275°F) for 3.5 minutes
Pre-vacuum Steam Sterilization
Full Cycle: Bagged -132°C (270°F) for 3 minutes
Alternate Method:
Place non-bagged instruments into the steam autoclave and
run at the listed cycles.
NOTE: Instruments sterilized unbagged should be used
immediately.
Drying:
To dry, use the drying cycle of the autoclave. Set cycle for
20-30 minutes.
Maintenance, inspection and testing:
• Visually inspect to ensure all contamination is removed.
ENGLISH
Содержание Cavitron
Страница 43: ...43 ITALIANO...
Страница 44: ...44 Cavitron Cavitron Cavitron Centers for Disease Control and Prevention CDC...
Страница 45: ...45 Cavitron 35 Cavitron Cavitron Cavitron Cavitron...
Страница 47: ...47 1 2 3 4 5 6 7 8 9 Cavitron...
Страница 48: ...48 15...
Страница 49: ...49 135 C 275 F 3 5 132 C 270 F 3 20 30...
Страница 50: ...50 1 c 2 3...
Страница 52: ...52 4 DENTSPLY DENTSPLY Cavitron DENTSPLY Professional 1 800 989 8826...
Страница 53: ...53...
Страница 54: ...54...
Страница 55: ...55...