6 Maintenance
6.2 Calibrating the scanner
67 97 216 D3801
D3801.201.01.03.02 2022-09
65
Calibrating the scanner
NOTE
Use calibration set only with clean, dry Primescan scanner
In order to achieve optimum results, the Primescan scanner must be
clean, disinfected and dry before the calibration.
➢ Make sure that the Primescan scanner is clean, disinfected and
dry.
1.
Remove the protective cap from the calibration set.
2.
Mount the calibration set on the tip of the scanner until it locks into
place.
3.
Secure the Primescan scanner in the calibration set using one
hand. Ensure that the external calibration set screw is fully screwed
in a clockwise motion until it gently locks into place.
4.
Click on the
"OK" button.
Ä
The measuring process starts.
Ä
The software prompts you to proceed to the next latching.
5.
Turn the screw counter-clockwise until you reach the next latching
point.
6.
Click on the
"OK" button. Hold the Primescan scanner still.
Ä
The software confirms the calibration process.
Ä
The software prompts you to proceed to the next latching.
7.
Complete steps 5 and 6 a total of 17 times.
Ä
The software provides information on the progress of the
calibration and informs you once the procedure is complete.
Ä
You will be prompted to measure the position of the exit
window.