6 Maintenance
6.1 Cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions Primescan Connect
60
67 97 216 D3801
D3801.201.01.03.02 2022-09
6.1.3.9
Using disposable sleeves
The Primescan scanner can be operated with disposable plastic
sleeves in order to ensure maximum infection control. The sleeves are
available using order number 66 86 880.
NOTE
The disposable sleeves cannot be used in the following
circumstances:
- when using color analysis.
1.
Following each patient, remove the disposable sleeve from the
scanner body and dispose of it according to the standard
procedure.
2.
Clean the scanner head with a dry cloth.
3.
Remove one disposable sleeve from its packaging. Slide the sleeve
up to the stop on the scanner body, so that it is securely positioned
on the Primescan scanner. To prevent the Primescan scanner from
falling, hold the Primescan scanner in other places, not just on the
sleeve.
4.
Position the Primescan scanner in the holder, so that it can warm
up briefly prior to the intraoral exposure. Make sure that the holder
has been wipe-disinfected (see "Non-critical surfaces [→ 48]").