4 Installation and commissioning
4.4 Initial startup
67 97 216 D3801
D3801.201.01.03.02 2022-09
33
4.4.2
Switching the units on
NOTE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
ü
Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is
absolutely dry (for at least one hour)
Ä
The unit is dry and can be put into operation.
Figure similar
1.
Switch on the notebook at the operating switch A.
2.
After loading the operating system, start the Connect SW
application by double-clicking on the Connect SW button.
NOTE
In order to prevent data security breaches, it is recommended that
users enable the password-protected login function for the Windows
operating system.