4 Installation and commissioning
4.4 Initial startup
Dentsply Sirona
Operating Instructions Primescan Connect
32
67 97 216 D3801
D3801.201.01.03.02 2022-09
A
B
C
D
E
F
G
H
I
1.
Place the window sleeve onto the scanner (A). Exercise a great
deal of care. Push the window sleeve carefully onto the cone until it
latches in.
2.
Carefully insert the connector of the scanner cable into the coupling
of the coupling box (B), watching out for the guide nose.
3.
Place the scanner in the scanner holder (C).
4.
Connect the coupling box to the laptop using a USB cable (D) with a
USB female connector.
5.
Insert the connector (E) of the supplied power supply unit (F) into
the coupling box.
6.
Connect the power supply unit to a supply network with protective
conductor using the power cable (G).
7.
Connect the laptop to a supply network with protective conductor
using the power supply unit (H) and the power cable (I).
8.
Check the plug connections to the mains connection and the
scanner. The scanner always remains connected.