6 Maintenance
6.1 Cleaning, disinfection, and sterilization
67 97 216 D3801
D3801.201.01.03.02 2022-09
51
6.1.3.2
Components of the scanner
CAUTION
Risk of cross-contamination
The scanner must not be used inside the patient's mouth without a
window sleeve/disposable sleeve/autoclavable sleeve. Without the
window sleeve/disposable sleeve/autoclavable sleeve, the scanner
cannot be disinfected or sterilized and cross-contamination may
occur.
➢ Always put on the window sleeve, disposable sleeve or
autoclavable sleeve with disposable windows.
CAUTION
Risk of injury
The window of the window sleeve is made of glass and is fragile.
➢ Use the sleeve with care so that the glass does not crack.
➢ If the glass has been cracked, the sleeve must no longer be used
on patients.
A
Area for gripping in order
to remove the sleeve from
the body.
C
Sleeve window
(made of coated sapphire for
window sleeve / made of plas-
tic for disposable sleeve and
autoclavable sleeve)
B
● Window sleeve
(standard sleeve)
● Disposable sleeve
● Autoclavable sleeve
with disposable
windows
D
Scanner window