10 Regular function test of the light barrier and the Start key
Dentsply Sirona
Operating Instructions inEos X5
50
64 17 096 D3586
D3586.201.01.11.02 07.2018
Function test 3
ü
The unit is switched on at the main switch.
ü
The power LED lights up green.
➢ Press and hold the Start key.
Operating state LED
Test result
The operating state LED changes
from green to red.
Nominal state
Stop function guaranteed.
The operating state LED stays
green or switches to yellow.
Error state
Call customer service.
No maintenance other than the light barrier function tests described
above must be performed on the inEos X5 .