10 PC Diagnostic Tool
10.3 Test procedure
Dentsply Sirona
Operating Instructions CEREC AC Connect
78
64 32 087 D3492
D3492.201.07.24.02 07.2018
Ä
The test starts. The entire test run takes approx. 30 minutes.
Ä
The result, i.e.
"Pass"
or
"Fail"
, appears at the end of the
test (see sections entitled "Test result: Pass" or "Test result:
Fail").
3.
On completion of the test, press any key to go on to the test dialog.
Tip: You can scroll to the individual test steps using the arrow keys.
The corresponding result is shown in front of each test step.
4.
To end the test, restart the PC.
Test result: Pass
No errors were found on the PC-specific hardware. Replacing the PC
component or the PC is not advisable.
1.
Perform the separate test for the supply board.
2.
Perform an image restore with the restore set for troubleshooting.
3.
Check the service instructions for other possible fault sources.