7 Operation
7.3 Taking acquisitions with the CEREC Omnicam
Dentsply Sirona
Operating Instructions CEREC AC Connect
42
64 32 087 D3492
D3492.201.07.24.02 07.2018
WARNING
Risk of injury for those diagnosed with epilepsy
For persons who have been diagnosed with epilepsy, there is a risk of
epileptic shock through the flashing light of the CEREC Omnicam.
➢ Patients who have been diagnosed with epilepsy cannot be
treated with the CEREC Omnicam.
➢ Dentists and dental assistants who have been diagnosed with
epilepsy cannot work with the CEREC Omnicam.
ü
The teeth are blow-dried.
1.
Change to the
"ACQUISITION" phase.
Ä
The camera is ready for scanning.
Ä
A live image appears which can be used to look around the
patient's mouth.
2.
Remove the CEREC Omnicam from its holder.
Ä
As soon as the camera is pointed over a tooth or the gums,
data acquisition begins. During the continuous data acquisition,
a color 3D model is generated automatically on the screen.
A white field indicates in which area data will be acquired. If the
automatic data flow breaks off, the white field is lost and the
audio signal changes. In this case, move the camera to any
area which has already been scanned. The scanning procedure
continues.
3.
Activate the foot control or point the camera cursor to the Omnicam
icon in the bottom left corner to end the acquisition procedure.
Proceeding with scanning procedure
1.
Activate the foot control or click on the Omnicam icon with the
cursor.
Ä
The scanning procedure begins.
2.
Proceed with the scanning procedure as described above.