![Dentsply Sirona CEREC AC Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/dentsply-sirona/cerec-ac/cerec-ac_operating-instructions-manual_2483624076.webp)
10 PC Diagnostic Tool
10.2 Start diagnostic tool
Dentsply Sirona
Operating Instructions CEREC AC Connect
76
64 32 087 D3492
D3492.201.07.24.02 07.2018
4.
Set the entry "
ASM1061 Storage Controller
" to "
IDE Mode
".
5.
In the menu go to "
Exit
" and "
Save Changes & Reset
" or press
the "F10" button to save the settings.
6.
Confirm the confirmation prompt with "
Yes
".
10.2
Start diagnostic tool
Test
Suitable for
Customer Diagnostics
Suitable for the user, in order to
check the PC components.
Technician Diagnostics
Suitable for the technician, in order to
check the PC components.
Sirona Windows Diagnostics
Windows Stress Test, in order to test
the PC at permanent load.
Tip: If one of the following steps does not work, further steps can be
found in the Service Manual.
ü
The PC is switched off.
1.
Switch on the PC.
2.
Repeatedly press the "F11" or "F8" button (in V2.2.1 - V2.2.3) until
the PC switches to the boot menu.
3.
Insert the diagnostics tool CD into the drive.
4.
Select the "
P3: TSSTcorp CDDVDW ...
" menu option.
5.
Confirm your selection with the
"Return"
key on the keyboard.
Ä
The diagnostics tool starts and a selection menu appears.
Tip: After one minute, if none of the arrow keys have been
pressed, the
"Customer Diagnostics"
test starts
automatically.
Содержание CEREC AC