Dentsply Sirona
Operating Instructions CEREC AC Connect
8 Maintenance
8.5 Charge battery (optional)
64 32 087 D3492
D3492.201.07.24.02 07.2018
71
8.5
Charge battery (optional)
NOTE
Reduced buffer cycles
After around 1000 buffer cycles the capacity of the battery fades due
to the nature of the battery technology used.
If the buffer times are too short, you should replace the battery.
The battery is permanently charged during operation on mains voltage.
This allows brief buffer operation after one hour of charging.
For a complete charge, the battery must be charged without interruption
for at least 12 hours. Keeping the acquisition unit connected to the
mains voltage and the power switch on is sufficient here. The PC does
not have to be switched on for the charging process.
NOTE
Reduced battery service life
If the battery is not charged over a long period of time, this
significantly reduces its service life.
➢ Always recharge the battery fully after buffer operation.