• If the pump is dropped or hit, the battery door may be-
come broken or damaged. Do not use the pump if the
battery door is damaged because the battery will not be
properly secured; this may result in loss of power or non-
delivery of drug.
• Prior to starting infusion, inspect the fluid path for kinks, a
closed clamp, or other upstream obstructions, and remove
all air to prevent air embolism.
• Close the tubing clamp before removing the cassette from
the pump to prevent unregulated gravity infusion.
• You must use a CADD
®
Extension Set with Anti-Siphon
Valve or a CADD
®
Administration Set with either an
integral or add on Anti-Siphon Valve to protect against
delivery inaccuracies and unregulated gravity infusion that
can result from an improperly attached cassette.
• For detailed instructions and warning pertaining to
Medication Cassette
™
Reservoirs or CADD
®
Administration
Sets, please refer to the instructions for use supplied with
those products.
• Frozen Medication must be thawed at room temperature
only. Do not heat the Medication Cassette
™
Reservoir in a
microwave oven as this may damage the medication, the
Medication Cassette
™
Reservoir, or cause leakage.
• Attach the cassette (the part of the Medication Cassette
™
Reservoir or CADD
®
Administration Set that attaches to
the pump) properly. An improperly attached or detached
cassette could result in unregulated gravity infusion or a
reflux of blood.