Reservoir Volume
is zero
NEXT to continue
Upstream Occlusion
Press STOP to stop
Press NEXT to restart
(Model 6101 only)
Upstream Occlusion
Press STOP to silence
(Model 6101 only)
A l a r m s a n d M e s s a g e s
When you see:
Take this action:
Fluid is not flowing from the
fluid container to the pump.
Check for a kink, a closed clamp,
or an air bubble in the tubing
between the fluid container and
pump. Press
⁄
to stop the
pump and silence the alarm for
2 minutes, then remove the
obstruction and press
„
to
restart the pump.
Fluid is not flowing from the
fluid container to the pump.
Check for a kink, a closed clamp,
or an air bubble in the tubing
between the fluid container and
pump. Press
⁄
to stop the
pump and silence the alarm,
then remove the obstruction
and press
„
to restart the
pump.
The Reservoir Volume has
reached 0.0 ml. Press
„
to
stop the alarm. Then install a
new fluid container, if appropri-
ate.