Deltec CADD-Prizm VIP CONTIN 6100 Скачать руководство пользователя страница 2

Table of Contents

Introduction .................................................................

1

Warnings ......................................................................

2

Cautions .........................................................................

4

CADD-Prizm

®

 Pump (Diagram) ...................................

6

Description of the Keys ...............................................

7

Installing a New Battery ..............................................

8

The Main Screen ........................................................... 10

The HELP Key 

Å

....................................................... 11

Starting the Pump ........................................................ 12

Stopping the Pump ...................................................... 13

Removing a Cassette .................................................... 14

Attaching a Cassette .................................................... 15

Resetting the Reservoir Volume ................................... 18

Priming the Tubing and Starting the Pump ............... 19

Inserting the Tubing into the Air Detector ................. 21

What if I drop or hit the pump? .................................. 22

Alarms and Messages .................................................. 23

Содержание CADD-Prizm VIP CONTIN 6100

Страница 1: ...31 Model 6100 and 6101 Ambulatory Infusion Pumps PAT I E N T I N F O R M AT I O N CONTIN Continuous Delivery Mode...

Страница 2: ...New Battery 8 The Main Screen 10 The HELP Key 11 Starting the Pump 12 Stopping the Pump 13 Removing a Cassette 14 Attaching a Cassette 15 Resetting the Reservoir Volume 18 Priming the Tubing and Star...

Страница 3: ...reprogrammed as your medication needs change The pump is programmed to deliver your medication con tinuously at a steady rate The pump stores programmed information and historical information that yo...

Страница 4: ...f under delivery will increase and could on occasion be significant You must regularly compare the volume remaining in the syringe to the pump s displayed values such as Reservoir Volume or Given to d...

Страница 5: ...integral or add on Anti Siphon Valve to protect against delivery inaccuracies and unregulated gravity infusion that can result from an improperly attached cassette For detailed instructions and warnin...

Страница 6: ...pump at temperatures below 2 C 36 F or above 40 C 104 F Do not store the pump at temperatures below 20 C 4 F or above 60 C 140 F Do not store the pump with a Medication Cassette Reservoir or CADD Admi...

Страница 7: ...5 Do not use the pump in the presence of flammable anes thetics or explosive gasses Use only Deltec accessories as using other brands may adversely affect the operation of the pump...

Страница 8: ...hes to the pump Battery Compartment CADD Prizm Pump Diagram Green Light Blinks every 3 seconds when the pump is running and deliver ing medication Amber Light Flashing pump is stopped or an alarm exis...

Страница 9: ...p Key explains what you see on the display Used by the clinician Advances you from one screen to the next and silences some alarms This key is not used in the Continuous Delivery mode Used by the clin...

Страница 10: ...ly which could result in death or serious injury Always have new batteries available for replacement If power is lost non delivery of drug will occur and depend ing on the drug being administered coul...

Страница 11: ...y The pump will beep if the battery is inserted correctly 5 Replace the battery door The pump will power up auto matically 6 Start the pump by pressing 7 When you see Start the pump press NOTE If you...

Страница 12: ...and the pump will beep at the end of the power up sequence All of the display indicators the software revision level and each parameter will appear briefly CAUTION Do not store the pump for prolonged...

Страница 13: ...r Volume screen you can press and the following will appear The symbol in the lower right corner means there are more help screens Press again to see the next help screen To page through all the help...

Страница 14: ...kinks a closed clamp or other upstream obstructions and remove all air bubbles to prevent air embolism An undetected upstream occlusion may result in under or non delivery of drug and depending on the...

Страница 15: ...ss NEXT to advance Stopping the Pump Stopping the pump stops delivery of medication Whenever the pump is stopped the amber light will blink 1 Press 2 When Stop the Pump ap pears press STOPPED will app...

Страница 16: ...top the pump appears press 2 Close all tubing clamps and discon nect the tubing from your access device as instructed by your clini cian 3 Use a coin to unlatch the cassette Insert the coin into the s...

Страница 17: ...led instructions and warnings pertaining to the Medication Cassette Reservoir or CADD Administration Set please refer to the instructions for use supplied with those products Frozen medication must be...

Страница 18: ...tach the cassette the part of the Medication Cassette Reservoir or CADD Administration Set that attaches to the pump properly An improperly attached or detached cassette could result in unregulated gr...

Страница 19: ...howing the type of cassette you have latched Press 7 Reset Reservoir Volume to may appear Press to reset the Reservoir Volume to the originally programmed value Admin set latched Next to continue Rese...

Страница 20: ...will be prompted to reset the Reservoir Volume You can however reset the Reservoir Volume without changing the cassette 1 Begin at the Main Screen Make sure STOPPED appears in the Main Screen 2 Press...

Страница 21: ...Make sure you are not connected to the pump s tubing Open the tubing clamps WARNING Do not prime the fluid path with the tubing connected to your catheter This could result in over delivery of medicat...

Страница 22: ...section If not follow your clinician s in structions for connecting the tubing to your access site Open all clamps 6 Press to start the pump Starting pump will appear The pump will automatically revi...

Страница 23: ...1 If your clinician has instructed you to use the Air Detector open the Air Detector door and thread the tubing through the groove 2 Close the door making sure the tubing does not get pinched or kink...

Страница 24: ...e the dot on the latch lines up with the solid dot on the pump Gently twist and pull on the cas sette to make sure it is still firmly attached Check the battery door to make sure it is still firmly at...

Страница 25: ...is action The 9 volt battery is low but the pump is operable Change the 9 volt battery soon Install a new 9 volt battery The pump will not run with a de pleted 9 volt battery A good battery must be in...

Страница 26: ...ress to stop the pump and silence the alarm for 2 minutes After you remove the cause of high pressure start the pump if necessary If the alarm continues contact your clinician Fluid is not flowing fro...

Страница 27: ...le in the tubing between the fluid container and pump Press to stop the pump and silence the alarm for 2 minutes then remove the obstruction and press to restart the pump Fluid is not flowing from the...

Страница 28: ...s When you see Take this action The Reservoir Volume value is low indicating that the level of fluid in the fluid container is low Prepare to install a new fluid container if appropriate There is a p...

Страница 29: ...27...

Страница 30: ...u c t i o n s Pump placement during bathing showering _____________ _______________________________________________________ Pump placement during sleep __________________________ _____________________...

Страница 31: ...29 N o t e s...

Страница 32: ...is a registered trademark of ULTRALIFE Batteries Inc These products are covered by one or more of the following U S Patent Nos 4 559 038 4 565 542 4 650 469 5 181 910 5 338 157 5 364 242 5 485 408 5 5...

Отзывы: