16
3. Place the pump upright on
a firm, flat surface. Press
down so the cassette fits
tightly against the pump.
4. Insert a coin into the latching
button, push in, and turn
counterclockwise until the
mark on the latch lines up
with the solid dot.
WARNING: Attach the cassette (the part of the Medication
Cassette
™
Reservoir or CADD
®
Administration Set that
attaches to the pump) properly. An improperly attached or
detached cassette could result in unregulated gravity infu-
sion of medication from the fluid container or a reflux of
blood, which could result in death or serious injury.
5. Gently twist and pull on the
cassette to make sure it is
firmly attached.
2000-03-08 D. Zurn
P i
Att
h C
T bl
®
®
2000 03 08 D Zurn
®