14
2000-03-08 D. Z
«Prizm Remove C
Stop the Pump?
Press Y or N
R e m o v i n g a Cassette
WARNING: Close the tubing clamp before removing the
cassette from the pump to prevent unregulated infusion,
which could result in death or serious injury.
To remove a cassette:
1. Press
⁄
to stop the
pump. When
Stop the
pump?
appears, press
´
.
2. Close all tubing clamps and discon-
nect the tubing from your access
device as instructed by your clini-
cian.
3. Use a coin to unlatch the cassette.
Insert the coin into the slot and
turn clockwise until the latching
button pops out.
NOTE:
If you cannot turn the
latching button, the cassette
may be locked. Contact your
clinician.
4. Remove the cassette from
the pump.
5. Discard the Medication
Cassette
™
Reservoir or
CADD
®
Administration Set
as instructed by your
clinician.
®
®