24
High Pressure
Pump Stopped
Delivery Stopped
(Model 6101 only)
Alarms and Messages
When you see:
Take this action:
There may be a kink in the
tubing or a clamp may be
closed. Unkink the tubing or
open the clamp and the pump
will resume delivery. You may
press
⁄
to stop the pump
and silence the alarm for
2 minutes. After you remove
the cause of high pressure, start
the pump if necessary. If the
alarm continues, contact your
clinician.
Fluid is not flowing from the
fluid container to the pump.
Check for a kink, a closed
clamp, or an air bubble in the
tubing between the fluid
container and pump. Press
⁄
to stop the pump and silence
the alarm for 2 minutes, then
remove the obstruction and
press
„
to restart the pump.