17
10.
TECHNICAL
CHARACTERISTICS
Instrument
Dimensions (Length x Width x Height)
140 x 88 x 38 mm
Weight
160g (complete with batteries)
Materials ABS
Display
2x4½ digits plus symbols
Visible area: 52 x 42 mm
Operating conditions
Operating temperature
-5…50°C
Warehouse temperature
-25…65°C
Working relative humidity
0…90%RH without condensation
Protection degree
IP67
Power
Batteries
3 1.5V type AA batteries
Autonomy
200 hours with 1800mAh alkaline
batteries
Power absorbed with instrument off
< 20
μ
A
Connections
Inputs for probes
2-pole female polarized standard
miniature connectors
Unit of Measurement
°C - °F;
Measurement of temperature by Instrument
Tc measurement range: K
-200 … +1370°C
Tc measurement range: J
-100 … +750°C
Tc measurement range: T
-200 … +400°C
Tc measurement range: E
-200 … +750°C
Resolution 0.1°C
Instrument accuracy
Thermocouple K
±0.1°C up to 600°C
±0.2°C over 600°C
Thermocouple J
±0.1°C up to 400°C
±0.2°C over 400°C
Thermocouple T
±0.1°C
Thermocouple E
±0.1°C up to 300°C
±0.2°C over 300°C
The precision only refers to the instrument. Error due to the thermocouple or to the cold
junction reference sensor is not included.
Temperature drift @20°C
0.02%/°C
Drift after 1 year
0.1°C/year
Содержание HD2328