9
3.
THE
PROGRAMMING
MENU
To access to the programming menu press simultaneously the
DATA+(A-B)
keys.
The items to be set are listed in this order:
1.
Selection of the unit of measurement:
the
"SEL_MEAS_UNIT"
message is displayed in
the comment line. The main line in the center of the display shows the selected unit of
measurement: Celsius (°C) or Fahrenheit (°F) degrees.
-
use the
and
arrows (respectively located above the HOLD and REL keys) to
modify
the type of probe;
-
press
DATA/ENTER
to
confirm
the modification and go onto the next item;
-
press
CLR
/
ESC
to
cancel
the modification;
-
to
exit
the menu, press the
A-B
/
MENU
key again.
2.
Selection of the type of Thermocouple:
the
"SEL"
message is displayed in the main line,
while the type of probe is displayed in the comment line; in the secondary line is shown the
"tc"
message.
The types of thermocouple that can be selected are: K, J, T, or E.
-
use the
and
arrows (respectively located above the HOLD and REL keys) to
modify
the type of thermocouple;
-
press
DATA/ENTER
to
confirm
the modification and go onto the next item;
-
press
CLR/ESC
to
cancel
the modification;
-
to
exit
the menu, press the
A-B/MENU
key again.
3.
Starting the User calibration procedure:
the
"CAL_MODE"
message is displayed in the
comment line, and
"FACt"
is displayed in the main line.
The instrument is provided with the factory ("
FACt
") calibration. It is also possible to
perform a "USER calibration"
(
"USEr"
)
of insprobe. The calibration information is
saved in the instrument's memory. The same correction is applied to any probe connected to
the input: therefore the "USER calibration" should only be used with the probe used for
calibration and not with other probes.
-
use the
and
arrows (respectively located above the HOLD and REL keys) and select
USEr
, to
access
the "User Calibration" procedure;
-
press
DATA/ENTER
to
confirm
the modification;
-
the
"SEL_CHAN"
message is displayed in the comment line;
-
use the
and
arrows (respectively located above the HOLD and REL keys) to select the
input
"A"
or
"B"
in the main line;
-
press
DATA/ENTER
to
confirm
the modification;
-
the
"SEL_MEAS_1/2"
message is displayed in the comment line;
-
use the
and
arrows (respectively located above the HOLD and REL keys) to select
"0",
"1"
or
"2"
in the main line;
-
press
DATA/ENTER
to
confirm
the modification;
-
press
CLR/ESC
to
cancel
the modification;
-
to
exit
the menu, press the
A-B/MENU
key again.
See chapter 4 for further explanations.
Содержание HD2328