13
6.
INSTRUMENT
SIGNALS
AND
FAULTS
The following table lists all error indications and information displayed by the instrument and
supplied to the user in different operating situations:
Display indications
Explanation
1ST_MEAS UP DOWN
Correct the first point using the arrows
/
2ND_MEAS UP DOWN
Correct the second point using the arrows
/
BATT TOO LOW
CHNG NOW
Indication of insufficient battery charge appearing on turning on. The
instrument issues a long beep and turns off. Replace the batteries.
BURN
The probe is disconnected or faulty.
CAL
LOST
Program error: it appears after turning on for a few seconds. Contact the
instrument's supplier.
CAL
FACT
Factory calibration.
CAL
USER
User calibration.
CAL_MODE
Calibration mode
ERR
This message appears when a wrong function is called: for example,
should the A-B key be pressed or the AVG function be enabled when one
of the two probes is not connected.
FUNC CLRD
Max, min and average values clearing
OVER
or
UNDR
Measurement overflow: indicates that the probe is measuring a value
exceeding the measuring range.
PLS_EXIT >>> FUNC
RES_FOR_FACT ONLY
Please exit using ESC >>> function reserved to factory calibration
SEL CHAN
Input channel selection
SEL MEAS 1/2
Select the first/second calibration point
SEL MEAS UNIT
Selection of the unit of measurement
SEL_TYPE_TC
Selection of the type of thermocouple
SYS
ERR
#
Instrument management program error. Contact the instrument's supplier
and communicate the numeric code # reported by the display.