Rev 1
–
Jul 2021
Page | 9
changes to the species/material selection field will NOT be saved
until the center button is pressed to confirm. When the species or
material is changed, all saved readings within the meter will be
deleted. If connected to the EDGE
™ app
, users will be prompted
to export readings before they are deleted from the app. Please
see Delmhorst E
DGE™
App User Guide for further instructions
regarding exporting readings and changing the available
species/material corrections in the BDX-30.
2.
Live Reading Area
Info: The live reading area displays the corrected moisture value of the
material. The reading is corrected using the Material Type (1),
Material Temperature (4), and Pin Correction (5) (see Figure 2
above).
Indicated readings with a less than (<) or greater than (>) sign are
considered Out of Range (OOR). OOR readings can be saved to
memory and exported but will not be used in statistical
calculations.
Use: Use the navigation buttons to move to the live reading area (entry
will be confirmed when a live reading appears on screen). Tip: If a
live reading is not currently being displayed, pressing the Read
button will navigate to the live reading area.
Take a Reading: Insert the pins into the desired material to their
full penetration (if possible). The moisture content of the material
will appear in the live reading area.
Hold a Reading: Press the Read button to hold the reading on
screen. HOLD will appear in the material selection line and the
meter will beep. A held reading can be saved, if desired (see
below). Saving a reading or pressing the Read button a second
time will return the meter to live reading mode.
www.
.com