Rev 1
–
Jul 2021
Page | 13
Table 2: Settings options
Setting
Description
Cal Check
•
Allows users to check the electrical calibration of the
meter
•
A value between 11.8 and 12.2 means the meter is
in calibration
•
A value of <11.8 or >12.2 means the meter is out of
calibration
–
change the batteries (2 x AA)
Bluetooth®
(BD-X30 only)
•
When Bluetooth is on but not connected, the
Bluetooth symbol will be on screen and flashing
•
When Bluetooth is on and connected, the Bluetooth
symbol will be on screen and solid
•
When Bluetooth is off, no symbol will be visible on
screen
•
Factory default is off
Temperature Unit
•
Changes the temperature unit between Fahrenheit
and Celsius
•
Factory default is Fahrenheit
Off Timer
•
Chose 1, 4, or 10-minute screen off timer
•
Factory default is 1 minute
Backlight
•
Turn backlight ON enable, and OFF to disable
•
Turn the backlight ON when in low ambient light,
and OFF when in bright ambient light
•
When set to AUTO, meter will automatically enable
and disable backlight according to ambient light
•
Factory default is off
Brightness
•
Adjust backlight brightness from 1 (low) to 10 (high)
•
The selected brightness level will be used whenever
backlight is enabled (ON or AUTO)
•
Factory default is brightness level 2
Contrast
•
Adjust the contrast level of the screen from 1 (low)
to 10 (high)
•
Factory default is contrast level 5
www.
.com
1.800.561.8187