
6. Troubleshooting
6.1 Alarms
6.1.1 Troubleshooting alarms
INFO
During commissioning, there is often a wider range of possible causes for alarms. This makes it difficult to provide
comprehensive alarm troubleshooting information. For example, during commissioning, the alarm may be activated
because of a wiring mistake. Careful thought, analysis and testing may be needed to find an alarm's cause.
See
Troubleshooting alarms
in the
Operator's manual
for troubleshooting information for a number of common
alarms.
6.2 Terminal damage
6.2.1 Troubleshooting terminal damage
All the terminals are tested before the controller leaves the factory. Therefore, the controller terminals do not generally need
to be checked, and problems with the terminals are rare.
However, the terminals can be damaged by electrostatic discharge, or by exposure to voltages and/or currents that are too
high. The hardware module self-check can be OK even though terminals are damaged.
If a high voltage or high current event, and/or the terminal input/output makes you suspect that the terminals are damaged,
then test the terminals.
If terminals are damaged, you can move the input/output to another set of terminals (if there are terminals available).
Remember to change the I/O configuration in PICUS accordingly.
Alternatively, replace the hardware module that has damaged terminals.
GPU 300 Commissioning guidelines 4189341033 UK
www.deif.com
Page 31 of 36