Page
28
of
32
For more information:
970.493.1901 or
Document #7650-0004 Rev C2. © Copyright 2019 Directed Energy, Inc. All rights reserved.
TRIG:EXT:TERM?
Returns the termination impedance of the external trigger input in ohms.
Example:
TRIG:EXT:TERM?
50 Ohms
TRIG:EXT:TERM n
Sets the termination impedance of the external trigger input in ohms. The
value of n must be 50 or 10000. Example:
TRIG:EXT:TERM 50
TRIG:EXT:TERM 50
TRIG:INTERNAL
Sets the trigger to internal. Example:
TRIG:INTERNAL
Internal
TRIG:INT:FREQ?
Returns the internal trigger frequency in hertz. Example:
TRIG:INT:FREQ?
2400 Hz
TRIG:INT:FREQ n
Sets the internal trigger frequency in hertz. The value of n must be from 40 to
100000. Example:
TRIG:INT:FREQ 2400
2400 Hz
TRIG:INT:MAIN:PULSEWIDTH?
Returns the pulse width of the main current pulse, in nanoseconds.
Example:
TRIG:INT:PULSEWIDTH?
280000 nsec
TRIG:INT:MAIN:PULSEWIDTH n
Sets the pulse width of the main current pulse in nanoseconds. The value of
n must be from 50 to 5000000. Example:
TRIG:INT:MAIN:PULSEWIDTH 280000