BIAS:CURRENT:SETPOINT?
Returns the amount of bias current in mA. Example:
BIAS:CURRENT:SETPOINT?
1500 mAmps
BIAS:CURRENT:SETPOINT n
Sets the amount of bias current in mA. The value of n must be between 0
and 5800. Example:
BIAS:CURRENT:SETPOINT 1500
BIAS:CURRENT:SETPOINT 1500
FAULT?
Returns the last 10 faults that have occurred since power-up. The codes for
the faults are:
0 = No Fault
1 = Keyswitch was turned off while the instrument was enabled
2 = Rear Enable was disabled while the instrument was enabled
4 = Ambient or exhaust temperature was out of range while the instrument was enabled
8 = AC-DC supply was out of range
16 = An open load was detected
The first listed fault is the most recent. The example below shows there have
been no faults since power-up:
FAULT?
0, 0, 0, 0, 0, 0
FAULT:CLEAR:ALL
Erases all previous faults. Example:
FAULT:CLEAR:ALL
All Faults Cleared
*IDN?
Short for Identity; returns the manufacturer’s name, unit model number,
serial number, firmware revisions, and manufacture date. Example:
*IDN?
DEI Scientific, PCX-7420, 00081, Q03 M04 C05, 00