Define Instruments PRO-WEI100 Скачать руководство пользователя страница 29

PRO-WEI100-MAN-19V01 (0225) 

Copyright © 2019 Define Instruments

29

A

APPENDIX A - INPUT FUNCTIONS

A number of user programmable functions are accessible from the 

 button and 

rear pins (see 6.2H–K). Note that availability of the 

TARE

/

BATCH

 and 

START

/

STOP

 

features are subject to your weighing mode, as selected in 6.2F.

User programmable input functions

Function Btn/Pin & Activation 

Time

Description

HOLD

Continuous

Freezes the display value.

LOCK

Continuous

Locks the control panel.

TARE
(see 2.3)

½ sec

Tares display value (flashes 

TARE

).

2+ sec

Resets tare to zero (flashes 

TR RST

 and then shows gross).

BATCH
(see 2.1)

½ sec

If no batch operation is active:

Performs a batch operation; display shows 

BATCH

If a batch operation is in progress:

Current batch is paused; display shows 

PAUSE

If a batch operation has been paused, or you have just 
powered up following a power loss during batching:

Controller will resume without resetting the batch 
value; display shows 

RESUME

2+ sec

Resets the batch value to zero and flashes 

BT RST

. Any 

current batching operations will be halted.

ZERO
(see 2.4)

½ sec

Zeroes the weight if the zero offset (i.e. the difference be-
tween the current no load weight and the calibration no 
load value) is within the 

ZERO BAND

 (see 2.4).

If the offset is less than the limit set in 

ZERO BAND

 (see 6.4E), 

the zero value is updated and the display flashes 

ZERO

. If the 

offset is greater than the limit set in 

ZERO BAND

, the zero 

value is not updated and the display scrolls 

OUTSIDE OF ZERO 

BAND!

2+ sec

Resets the zero offset value to the original calibration off-
set value. Display flashes 

RS ZER

 and then shows gross.

Содержание PRO-WEI100

Страница 1: ...ing to analog serial in puts or PCs PLCs HMIs Contents 1 Specifications 2 2 Features 3 3 Front Panel Display 6 4 Wiring 8 5 Dimensions Installation 11 6 Input Setup Calibration 13 7 Setpoint Setup 24...

Страница 2: ...20mA 0 10V Comm Port Number of comm ports None or 1 Comm port options S2R Isolated RS232 RJ terminal or S4S Isolated RS485 screw terminal Serial output Custom ASCII Modbus RTU slave Gedge Ranger A or...

Страница 3: ...in in Weight GIW Batching Direction The Batching Direction parameter is set in 6 2G and should be set to GIW Gain in Weight for applications where the weight increases as product is added to the weigh...

Страница 4: ...changed it is possible to configure SP 2 so that it tracks SP 1 and always turns off at a fixed amount below the required batch weight E g If your initial batch weight was 100Kgs and you wanted SP 2 t...

Страница 5: ...lue If the input value on the load cell is not within the Capture Band setting then the controller displays the current loadcell value and does not zero Capture Band can be set from 1 to 254 counts an...

Страница 6: ...see 6 2H This button is typically used to scroll through options or increase values in the setup menu Pressing this button from the main display will allow you to view reset the Peak value and view t...

Страница 7: ...table below and will cycle continuously at each press of the or button Press at any time to return to normal operating mode PEAK and VALLEY may be reset to zero by pressing the and buttons at the sam...

Страница 8: ...Analog Output See 4 4 4 1D Analog Input See 4 2 4 1E Function Pins See 4 6 4 1F Power Supply HV LV See 4 7 4 2 Wire the strain gauge input module Wire your loadcell input module as shown in the dia g...

Страница 9: ...ntroller See 4 1A SP2 SP1 R2 SP3 SP4 SP1 SP2 R4 4 4 Wire the analog output if installed If your controller has analog output fitted wire it as shown for either voltage 0 10V or current 4 20mA See 4 1C...

Страница 10: ...he power is on NEVER connect your low voltage controller to mains power Wire your controller for low or high voltage power supply as show in the diagrams below Check the label on the unit against the...

Страница 11: ...NSIONS INSTALLATION 5 1 Case dimensions Cabling Allowance 96mm 3 78 92mm 3 62 120mm 4 72 39mm 1 54 155mm 6 10 minimum depth required behind panel 4mm 0 16 Cabling Allowance 48mm 1 89 45mm 1 77 120mm 4...

Страница 12: ...ter back C Slide the Panel Gasket over the rear of the unit to the back of the Meter Faceplate D From the front of the panel insert the meter into the Panel Cutout Holding the unit in place engage the...

Страница 13: ...scrolls across the display and toggles with SKIP Press to skip to 6 3 or the button and then to ENTER input setup B _ _ _ SUPPLY REJECTION FREQUENCY scrolls across the display Use the and buttons to s...

Страница 14: ...mptying operations or GIW gain in weight for filling operations Press See Section 2 1 for additional information For 6 2H K please refer to the list of input functions in Appendix A H _ _ _ PROG BUTTO...

Страница 15: ...libration This unit has been calibrated for 0 10 000 counts 2mV V sensor gain at full scale A _ _ _ CALIBRATION TECHNIQUE scrolls across the display and toggles with SKIP Press to skip to 6 4 or use t...

Страница 16: ...o adjust the no load value if required and then press D _ _ _ CALIBRATE SPAN scrolls across the display and toggles with the current selection Use the and buttons to select YES or NO and then press If...

Страница 17: ...be directed back to the operational display To enter step 6 4 you must select SKIP at 6 3A Zero offset calibration J _ _ _ ADJUST OFFSET scrolls across the display and toggles with the currently sele...

Страница 18: ...and then press 6 4 Zero maintenance See Section 2 4 for more information on zero maintenance parameters A _ _ _ ZERO MAINTENANCE scrolls across the display and toggles with SKIP Press to skip to 6 5 o...

Страница 19: ...easurement If the change in input exceeds the averaging window value it will not average ensuring fast response when there are large differences between readings E g When product is being dropped into...

Страница 20: ...output data source Use the and buttons to select an option from NET BATCH or LIVE and then press Note that where NET BATCH is indicated the option that will be displayed is controlled by the weighing...

Страница 21: ...s Please skip to 6 7G now See Appendix B 3 for more information on Gedge output formats D _ _ _ DATA SOURCE scrolls across the display and toggles with the currently selected serial data source Use th...

Страница 22: ...scrolls across the display and toggles with the currently selected parity Using the and buttons select NONE ODD or EVEN and then press J The step that you proceed to now will depend on the Serial Mode...

Страница 23: ...imer will be reset to zero and will immediately begin counting D _ _ _ DATE scrolls across the display and toggles with the current selection Use the and buttons to adjust the date from 1 to 31 and pr...

Страница 24: ...to enter your security code factory default 1 Then press If the correct PIN is entered setup is started at 7 2 If an incorrect PIN number is entered _ _ _ ACCESS DENIED scrolls across the display and...

Страница 25: ...display and toggles with the activation source for the selected setpoint Use the and buttons to choose NET BATCH or LIVE and then press Note that where NET BATCH is indicated the option that will be d...

Страница 26: ...K _ _ _ USER ACCESS scrolls across the display and toggles with the direct access permission setting for the selected setpoint Use the and to select either OFF or ON and then press When enabled this...

Страница 27: ...Pressing for the last ena bled setpoint will exit and return to the operational display 9 RESET PIN NUMBERS VIEW FIRMWARE VERSION If you have forgotten your PIN number s follow the procedure below to...

Страница 28: ...the current selection Use the and buttons to select YES or NO and then press If you selected YES connect a mA or volt meter across the analog output connector see 4 4 and then continue to 10D If you...

Страница 29: ...ion is in progress Current batch is paused display shows PAUSE If a batch operation has been paused or you have just powered up following a power loss during batching Controller will resume without re...

Страница 30: ...us Sets the peak value to the current input value flashes PK RST PRINT see B 5 Continuous Sends a single Ranger A output string SERIAL MODE 6 7B must be set to RNGR A and OUTPUT MODE 6 7G must be set...

Страница 31: ...r aborts the operation In Custom ASCII mode data is normally read as formatted data which includes decimals and any text characters that may be selected to show units However it is also possible to re...

Страница 32: ...he controller has completed a read or write instruction it responds by sending a carriage return line feed CR LF back to the host If the instruction was a read command the CR LF follows the last chara...

Страница 33: ...40066 SP3 40067 SP4 40068 40071 40074 Make delay SP1 40071 SP2 40072 SP3 40073 SP4 40074 32 Bit Signed 2 x 16 Bit LSW MSW Function 40515 40516 Tared Batch weight net 40517 40518 mV 40521 40522 Live w...

Страница 34: ...zero ASCII Space Stored tare is zero D7 P Print key operation ASCII Space No print key operation B 4 Ranger A mode Ranger A is a continuous output used to drive remote displays and other instru ments...

Страница 35: ...s set in 6 7D and the units selected in 6 7F Note that changing the display units in 6 7F does not perform any conversion calculations You will need to scale the instrument to match the printed units...

Страница 36: ...245 Westpark Village Auckland 0661 New Zealand Ph 64 9 835 1550 Fax 64 9 835 1250 sales defineinstruments co nz www defineinstruments com United States Dallas TX Ph 214 926 4950 sales defineinstrument...

Отзывы: