Otherwise, press the
button and then
to
ENTER
serial setup.
B _ _ _ SERIAL MODE
scrolls across the display and toggles with the current serial
mode. Use the
and
buttons to choose between:
ASCII
(custom),
MOD-
BUS
(RTU),
GEDGE
,
RNGR A
(Ranger A), or
. Then press
.
¨
If you selected
GEDGE
, continue to 6.7C now.
¨
If you selected
RNGR A
or
, skip to 6.7D now.
¨
If you selected
ASCII
or
MODBUS
, skip to 6.7H now.
See Appendix B for more information about the available serial modes.
C _ _ _ OUTPUT FORMAT
scrolls across the display and toggles with the currently
selected Gedge output format. Use the
and
buttons to choose between
C1
,
C2
, or
C3
, and then press
.
¨
Please skip to 6.7G now.
See Appendix B.3 for more information on Gedge output formats.
D _ _ _ DATA SOURCE
scrolls across the display and toggles with the currently
selected serial data source. Use the
and
buttons to select an option from:
NET
/
BATCH
,
LIVE
, or
PEAK
, and then press
.
Note that where
NET/BATCH
is indicated, the option that will be displayed is controlled by the
weighing mode selected in 6.2F.
NORMAL
mode =
NET
,
BATCH
mode =
BATCH
.
E
The step that you proceed to now will depend on the Serial Mode that you
selected in 6.7B:
¨
If your Serial Mode =
, continue to 6.7F now.
¨
If your Serial Mode =
RNGR A
, skip to 6.7G now.
F
_ _ _ PRINT UNITS
scrolls across the display and toggles with the current units
that will be printed on the
Weigh Ticket
(see B.5) when the print function is trig-
gered. Use the
and
buttons to choose between
NONE
,
GRAMS
,
KGS
,
TONNES
,
LBS
or
KN
, and then press
.
¨
Please skip to 6.7H now.