PRO-WEI100-MAN-19V01 (0225)
Copyright © 2019 Define Instruments
17
mV/V calibration
G _ _ _ ENTER TOTAL FULL SCALE WEIGHT OF LOAD CELLS IN COUNTS
scrolls
across the display and toggles with the current selection. Using the
and
buttons, enter the total full scale weight of the connected load cell(s) in counts,
referring to the load cell manufacturer's test certificate. Then press
.
H _ _ _ ENTER MV/V FROM LOAD CELL
scrolls across the display and toggles with
the current selection. Using the
and
buttons, enter the mV/V (or average
mV/V) of the connected load cell(s), and then press
.
I
_ _ _ SET ZERO NOW?
scrolls across the display and toggles with the current
selection. Use the
and
buttons to select
YES
or
NO
, and then press
.
Selecting
YES
sets your zero position when the load cells are powered up and in position.
mV/V calibration is now complete. You will be directed back to the operational
display. (To enter step 6.4, you must select
SKIP
at 6.3A.)
Zero (offset) calibration
J
_ _ _ ADJUST OFFSET
scrolls across the display and toggles with the currently
selected zero value. Place a known weight on the weigh platform if required,
and use the
and
buttons to enter the desired value. Press
to accept.
The offset will be automatically calculated to match the desired weight, and the scale factor will
not be altered. (Normally the weight would be removed and the value would be zero.)
Zero calibration is now complete. You will be directed back to the operational
display. (To enter step 6.4, you must select
SKIP
at 6.3A.)
E_Cal calibration
K _ _ _ E_CAL ZERO OFFSET
scrolls across and toggles with the current selection.
Use the
and
buttons to adjust the zero offset value, and then press
.
This is updated after each calibration to show the internal
ZERO OFFSET
E Cal value.
L
_ _ _ E_CAL SCALE FACTOR
scrolls across and toggles with the current selec-
tion. Use the
and
buttons to adjust the scale factor value, and press
.