PRO-WEI100-MAN-19V01 (0225)
Copyright © 2019 Define Instruments
13
6
INPUT SETUP & CALIBRATION
6.1 - Enter F1 PIN number
A
Enter the calibration mode by pressing the
button.
_ _ _ ENTER F1 PIN
scrolls across the display and toggles with
0
. Use the
and
buttons to enter your security code (factory default 1). Then press
. If the
correct PIN is entered, setup is started at 6.2.
If an incorrect PIN number is entered,
_ _ _ ACCESS DENIED
scrolls across the
display and it returns to normal operating mode.
You will have the opportunity to change your PIN number at the end of this section
(6.9). If you have forgotten your PIN number, see Section 9.
6.2 - Input setup
A _ _ _ INPUT SETUP
scrolls across the display and toggles with
SKIP
. Press
to
skip to 6.3, or the
button and then
to
ENTER
input setup.
B _ _ _ SUPPLY REJECTION FREQUENCY
scrolls across the display. Use the
and
buttons to select
50HZ
or
60HZ
, and then press
.
C _ _ _ SAMPLING RATE
scrolls across the display and toggles with the current
selection. Use the
and
buttons to select an option from the following list:
1HZ
,
2HZ
,
5HZ
,
10HZ
, or
50HZ/60HZ
(50 or 60Hz will depend on your selection
in 6.2B above). Then press
.
D _ _ _ DECIMAL POINT POSITION
scrolls across the display and toggles with the
current selection. Use the
and
buttons to select
NO DP
(default),
0.0
,
0.00
,
0.000
or
0.0000
, and then press
to accept and continue.
E
_ _ _ ROUNDING
scrolls across the display and toggles with the current display
rounding. Using the
and
buttons, select:
NONE
(default),
2
,
5
,
10
,
20
,
50
,
100
,
200
,
500
, or
1000
. Then press
.