4. Attach the Compression Module to the Frame.
Continue manual CPR.
Check the Compression Module to be sure the Patient Interface Pad and Battery Pack
are installed.
Insert the Module into the Frame and rotate in either direction until the Module is in line
with the Frame and snaps to lock into place.
5. Adjust the placement of the Compression Piston relative to the patient’s chest.
Adjust the Frame and Backboard to position the Compression Piston over the patient’s chest
and directly in line with the nipples. (This is the same target point used for manual CPR.)
WARNING
Do not initiate Lifeline ARM compressions if the piston is not in the proper position.
6. Turn the unit On.
Press the On/Off button on the Control Panel for at least 1 second. If the
Battery Pack indicator shows red for low battery or the device does not
turn on, replace the Battery Pack or connect the AC Adapter.
2. G
ro
up T
ra
in
in
g S
es
sion
s
RAC-A1515EN-BA
25
Содержание Lifeline ARM
Страница 2: ......
Страница 4: ...RAC A1515EN BA 4 ...
Страница 6: ...RAC A1515EN BA 6 ...
Страница 10: ...RAC A1515EN BA 10 ...
Страница 32: ...RAC A1515EN BA 32 ...
Страница 38: ...RAC A1515EN BA 38 ...
Страница 40: ...RAC A1515EN BA 40 ...
Страница 42: ...RAC A1515EN BA 42 ...
Страница 43: ......
Страница 44: ......