The Battery Pack
The Lifeline ARM is powered by a replaceable Battery Pack that slides into either side of the
Compression Module. The Battery Pack must always be installed in the unit to operate the
device, even when powered by the AC Adapter.
The Compression Module should be turned off, or paused if in use, whenever batteries are
swapped out. To remove the Battery Pack, squeeze the eject release latches on either side
of the Battery Pack opening.
To insert the Battery Pack, be sure the contacts are facing the device and push in until the
latch clicks. When the device is turned on, the Battery Pack Status indicator will display
throughout its use.
When fully charged, the Battery Pack will provide about 60 minutes of compressions. With
the Battery in the Compression Module at room temperature and in the off state, the external
AC Adapter can charge the battery in less than 3 hours.
To charge the Battery Pack with the AC Adapter:
1.
Insert the Battery Pack into the Compression Module.
2.
Connect the AC Adapter plug to the external power input
jack on the Compression Module. Lock it in place by aligning
the raised notch on the plug with the notch on the jack.
While the Battery Pack is charging, a green LED will flash
slowly on the Battery Pack Indicator.
To check the Battery Pack’s charge with LED status indicator:
Press and hold the membrane switch panel
on the Battery Pack. A full charge is indicated
by all-lit LEDs on the status indicator.
2. G
ro
up T
ra
in
in
g S
es
sion
s
RAC-A1515EN-BA
19
Содержание Lifeline ARM
Страница 2: ......
Страница 4: ...RAC A1515EN BA 4 ...
Страница 6: ...RAC A1515EN BA 6 ...
Страница 10: ...RAC A1515EN BA 10 ...
Страница 32: ...RAC A1515EN BA 32 ...
Страница 38: ...RAC A1515EN BA 38 ...
Страница 40: ...RAC A1515EN BA 40 ...
Страница 42: ...RAC A1515EN BA 42 ...
Страница 43: ......
Страница 44: ......