Page: 20 of 102
4.4
OMD-2008 Piping
Connect the OMD-2008 Monitor to the sample point of the oily-water separator outlet
and to a source of oil free water employing 6 mm OD copper or stainless-steel pipe. The
sampling point should be located on a vertical section of the separator outflow piping to
minimize the effects of any entrained air. The sampling point should be at a level above
the outlet of the monitor to ensure a proper sample flow rate.
Figure 4. Piping OMD-2008
Содержание OMD-2008
Страница 28: ...Page 27 of 102 5 5 Cabling OMD 2008 EV Figure 9 Bottom View OMD 2008 EV ...
Страница 84: ...Page 83 of 102 ...
Страница 85: ...Page 84 of 102 ...
Страница 86: ...Page 85 of 102 ...
Страница 87: ...Page 86 of 102 ...
Страница 88: ...Page 87 of 102 ...
Страница 89: ...Page 88 of 102 ...
Страница 90: ...Page 89 of 102 ...
Страница 91: ...Page 90 of 102 ...
Страница 92: ...Page 91 of 102 ...
Страница 93: ...Page 92 of 102 ...
Страница 94: ...Page 93 of 102 ...
Страница 95: ...Page 94 of 102 ...
Страница 99: ...Page 98 of 102 Type Approved by ...